Neomedica’s giant – a sophisticated solution for high-end hospitals, laboratories and clinicals with huge workload. State-of-the-art technology that meets all of the expectations, no matter how great they are
Phoenix NCC-5500
Description
Phoenix NCC-5500 is the most advanced 5-diff hematology analyzer in Neomedica’s portfolio with an autoloader with 120 positions equipped with a special intelligent recognition test tube bar coded rack which allows continuous sampling, as well as a rotary shear valve and emergency sample testing. This instrument is also able to perform automatic mixing of the samples and rinsing of the residual blood on the sample aspiration probe, where the contact with the samples is avoided and maximal security for the operator is thus insured. As the model with the highest throughput, it is intended for facilities with a big number of samples per day that gain the highest benefit and cost savings from automation of the process. With long-standing references from numerous top of the range hospitals and laboratories, this model continues to amaze the users with its stability, brilliant performances and exquisite appearance.
Reagents, consumables and controls
Reagents
N117145 | Neo-U-Diluent | 20 L |
N117325 | Neo Cleanser | 20 L |
N117125 | Neo-U-Sheath | 20 L |
N117212 | Neo-Lyser-WHP | 1 L |
N114100 | Neo-Probe Cleaner | 100 mL |
N117222 | Neo-U-Stain | 4 mL |
Phoenix NCC-5500
03K111TA3 | CBC-3K 3 levels (1+1+1) | 3×2.0 mL |
0CPO1OVA2 | CBC-Cal Plus | 1×2.0 mL |
Technical specification
Principle | WBC/DIFF: Flow cytometry, semiconductor, laser light multi-dimensional cell classification WBC Analysis: Optical and impedance measurementsRBC/PLT Analysis: Impedance MethodHGB Test: Cyanide-free reagent colorimetry |
Working mode | Two units (WOC & WIC) for selection of WBC, RBC, HGB, PLT and other elements and parameters |
5-part differentiation of WBC
28 parameters + 2 histograms for RBC and PLT, 2 5-part differential scatter-grams for eosinophils and neutrophils and 2 3-D stereograms |
|
Parameters | WBC, LYM#, MON#, NEU#, EOS#, BAS#, LYM%, MON%, NEU%, EOS%, BAS% , RBC, HGB, HCT, MCV, MCH, MCHC, RDW-SD, RDW-CV, PLT, MPV, PDW, PCT, P-LCR, P-LCC
3 RETIC parameters: RETIC, RETIC_ABS, IRF |
Mode | Whole blood, Prediluted |
Sample volume | Autoloader 50 µl, manual (emergency) 40 µl, pre-diluted mode 40 µl |
Dimension of aperture | WBC: 100 µm, RBC/PLT: 68 µm |
General throughput | Autoloader: up to 110 samples per hour
Manual mode: up to 100 samples per hour |
Storage | More than 200.000 sample results including histograms |
QC mode | LJ, X, XR, XB etc. |
Calibration mode | 3 calibration modes: automatic calibration, whole blood calibration and manual calibration |
Windows operating system | |
Display | PC display (14 inch color TFT) |
Power supply | AC100-240V~50/60 Hz |
Interface | PC display with Windows, bidirectional LIS and barcode reading |
Report | External printer |
Automatic reagent residual alarm | |
Dimensions | 670x790x690mm (WxDxH), 91,5 kg |