Automatic Bench Top Biochemistry Analyzer performing up to 200 tests/hour. Enhanced possibilities for veterinary practitioners.
NeoChem 100 VET
Description
With a brand new design, NeoChem 100 VET provides an excellent choice for small and medium-sized laboratories who would like to be equipped with an automated device. Small in size, but great in performance, this bench-top, fully automated, compact, biochemistry analyzer is equipped with a hot water 7-Step reaction cuvette washing system, on-board reagent cooling system and advanced photometric optical system with fiber optic. With 40 sample positions and random access feature as well as dedicated veterinary software, NeoChem 100 VET provides a comprehensive solution tailored to meet the needs of your laboratory. Upon request, the completely closed system can be provided for you, along with the high-quality bar-coded biochemistry reagents at most competitive prices.
Consumables
N1177500 | Detergent A | R1: 2×250 ml |
N1177501 | Neo-Wash-concentrate | R1: 2×100 ml; R2: 2x100ml |
* | Reagent bottle 5 ml | * |
* | Reagent bottle 20 ml | * |
* | Sample cups | * |
* | Reaction Cuvette | * |
NC100-200006 | Halogen lamp | 12V / 20W |
Technical specification
Testing speed | Up to 200 tests/hour ( measuring cycle 18 s ) |
Assay method | End Point (reagent or sample blank), Kinetic (Rate A, Rate B), Two Point |
Calibration mode | Single point, Two point or Multiple point, Logit Log, Segmentline, Exponential 5P, Polynomial, Poly4, Spline Calibration Curve. |
Reagent tray | 40 barcoded position for 5 ml and 20 ml bottles on removable tray |
Reagent volume | R1: 150 – 300 µl, R2: 10 – 150 µl in 1 µl step |
Sample tray | 40 positions, samples can be placed randomly, including standard, QC and STAT samples in tubes or sample cups |
Sample volume | range 2 – 50 µl in 0,1 µl step |
Reaction tray | 60 reusable plastic cuvettes in 6 segments (5 mm optical path) |
Reaction volume | range 150 – 300 µl |
Reaction temperature | 37 ± 0,1 ℃ |
Reagent temperature | From 4 – 8 ℃ |
Photometer | 9 wavelengths: 340, 405, 450, 510, 546, 578, 620, 660, 690 nm Static fiber optic system with multi wavelength reading |
Abs linearity | from 0,0000 up to 4,0000 Abs |
Wavelength accuracy | ± 2 nm |
Light source | 12V / 20W long life halogen lamp |
Interface | RS232 (instrument-PC), LIS supported |
Reagent detecting | Integrated bar code reader for reagents |
Alarms | Alarm for “Distilled water empty” and “Waste full” |
QC Rules | Multi QC function. For every test up to 3 different QC materijal |
Other specification | On-board mixer, Low water consumption, Automatic cuvette detection and selection, Minimal maintenance |
Dimensions | 415 x 565 x 385 mm (WxDxH), weight 28 kg |